Spectrophotometric Determination of Carvedilol in Bulk drug and its Formulation by Multivariate Calibration Technique

 

Anant Nayabaniya, Seetharaman R.*, Lakshmi K. S.

Department of Pharmaceutical Analysis, SRM College of Pharmacy, SRM Institute of Science and Technology, Kattankulathur – 603203, Kancheepuram, Tamil Nadu, India.

*Corresponding Author E-mail: anantnayabaniya@gmail.com, seerampharm@gmail.com

 

ABSTRACT:

The determination of Carvedilol in bulk as well as in formulations was done in this study by brining upon a new approach and validating it to be a sensitive, accurate and simple analytical method. The analytical approach was done using a UV spectrophotometer with the application of multivariate calibration. The relationship between concentration and absorbance taken at five different wavelengths including the absorption maxima has been used to develop this multivariate calibration technique. The linear regression equations, residual sum of squares and other mathematical as well as statistical approaches were used for the validation of process according to ICH Q2 (R1) guidelines. The validation parameters that need to be accomplished were precision, accuracy, linearity, and others as mentioned in the guidelines. There was linearity in the plot between the concentration and absorbance in the concentration level of 3 -5μg/mL with a good correlation coefficient of R2 = 0.9989. The linear regression equation calculated was: y = 0.1167x + 0.01944. The detection limit and the quantitation limit as calculated were 0.036μg/mL and 0.109 μg/mL respectively. This method has good sensitivity with satisfactory results and can be implemented in routine analysis of bulk drugs and formulations at low cost.

 

KEYWORDS: Beta-blocker, Carvedilol, Multivariate;calibration, UV Spectrophotometer, Validation.

 

 


INTRODUCTION:

Carvedilol [Figure 1] is a cardiovascular agent. It is a stimulator of electrophysiological properties of the heart which is done by combinatory alterations in Ca2+ and K+ ion channels. More precisely, carvedilol is an antagonist of α1 and β1, β2 receptors1,2,3. It also acts as an anti-oxidant to decrease the oxidative stress which prevents the mitochondria from oxidative;phosphorylation. There are many cardiovascular disorders like congestive heart failure, high blood pressure, myocardial infraction (MI) and ischemic heart disease (IHD) in which carvedilol is used in varying doses: 3.125mg. 6.25mg and 12.5mg. Carvedilol is available in tablet formulations under branded and generic name.

 

This drug is actually synthesized to contain the racemic mixture of dextro R[+] and leavo S[−] carvedilol because either of these forms doesn’t implicate pharmacological response as of the mixture. Chemically carvidol is described as [3-(9H-carbazol-4-yloxy)-2-hydroxypropyl] [2-(2-methoxyphenoxy) ethyl] amine, with a molecular formula of C24H26N2O4 and molecular mass of 406.474g/mol4. The solubility of Carvedilol in water and phosphate buffer (pH 6.6) was found to be 17.2±0.98g/100ml and 15.46±0.45g/100ml, respectively5. The FDA first approved carvedilol in 1995 for use in mild to moderate high blood pressure and angina pectoris6,7.

 

Alternatively carvedilol is a non-selective β -blocker commonly used in combination with other drugs8,9. The non-selective β -blocking action is shown by (S)-carvedilol, whereas the α -blocking activity is shown by the combination of R[+] and S[−] enantiomers10,11. Carvedilol is a third generation medicine of current interest with antihypertensive and vasodilating activities. of particular concern, carvedilol has been recognized as having novel antioxidant properties that may contribute to its therapeutic effects12. Carvedilol is also found to possess significant anti denaturing and anti-cataract activity. The anti-denaturing activity of the Carvedilol is related to the inhibition of thermally induced protein denaturation13.

 

The review of literature evidences methods like UV-Visible Spectrophotometry14-21, Spectrofluorometry 22, High Performance Liquid Chromatography (HPLC) 23-29, Capillary Electrophoresis30,31, High Performance Thin Layer Chromatography (HPTLC)32, etc., either as a single entity or with other drugs in pharmaceutical formulations or biological fluids. The purpose of this study was to improve the validated method with UV spectrophotometer for the determination of Carvedilol in the pure and tablet dosage forms according to ICH guidelines. The main purpose of this work was to establish a sensitive, cheap and non-destructive method for the analysis of carvedilol. Linear regression equations have been used to establish the relationship between the concentration and absorbance of the analyte at five different wavelengths33.

 

Multivariate calibration is a technique in which there is transformation of analysis of an independent variable from a single dependent variable to 'n' number of dependent variables. E.g. multi-wavelength spectral response to analyte concentration, molecular descriptors to biological activity, etc. This technique results in fast, cheap and non-destructive analysis. Multivariate;calibration collaborates with the statistical approach to obtain a significant resolving power, sensitivity, and cost optimization for the analysis of drugs and excipients whether it be a quantitative analysis or a routine analysis34-37. The models were tried to develop with varying Dλ. The best results can be obtained with absorbance at certain wavelengths intervals38. This approach is based on the reduction of multi-linear regression functions to uni-variate data set, which provides more sensitive determination than the classical UV method39. Validation has been proven assurance for the process efficiency and sturdiness and it is the full-fledged quality attributing tool for the pharmaceutical industries40.

 

Let us consider the spectrum of an analyte ‘x’, carvedilol in our case gives an absorbance ‘Aλ’ at five corresponding wavelengths (λ = 238, 240, 242, 244 and 246 nm), the Beer- Lambert’s equation for each selected wavelength will be:

Aλ238 = f x Cx + k1 ………………………………….. (1)

Aλ240 = g x Cx + k2 ………………………………….. (2)

Aλ242 = h x Cx + k3 ………………………………..… (3)

Aλ244 = i x Cx + k4 …………………………………... (4)

Aλ246 = j x Cx + k5 ………………………………….. (5),

where, Aλ → absorbance of the analyte,

f, g, h, i, j → slopes of the linear regression functions,

k1, k2, k3, k4, k5 → intercepts of the linear regression functions and

Cx → concentration of the analyte.

 

The above five equation systems (1-5) can be written as:

AT = f x Cx + g x Cx + h x Cx + i x Cx + j x Cx + kT ....(6)

Equation (6), can be further simplified to:

 

AT = Cx ( f+g+h+i+j ) + kT ...………………………...(7)

where, AT → sum of the absorbance obtained and KT → sum of intercepts of regression equations.

 

Therefore, the concentration of the analyte ‘x’ in a solution can be calculated by using the equation:

 

          (AT – kT)

Cx =  –––––––––––………………………………….. (8)

          (f+g+h+i+j)

 

MATERIALS AND METHODS:

Solvents and chemicals:

·       Methanol

·       Distilled water

·       Carvedilol of analytical grade was received from Ideal Analytical Research Lab, Pondicherry. The marketed tablet formulation used was Cardivas (label claim – 6.25mg), Sun Pharma, India, obtained from a local pharmacy.

 

Solubility:

·       Water: Practically insoluble

·       Dimethylsulfoxide: Freely soluble

·       Methanol: Soluble

·       Ethanol: Sparingly soluble

·       Ethyl ether: Slightly soluble

 

Instruments:

·       Lab India UV 3092 double beam spectrophotometer

·       Sonicator

·       Digital balance (Shimadzu AY220)

 

METHOD DEVELOPMENT:

Solvent selection:

While checking the solubility of Carvedilol, it was found to be freely soluble in methanol with slight or no solubility in other solvents. Therefore, carvedilol, both standard drug and the formulation were solubilized in methanol yo obtain a solution of various concentration for analysis.

Preparation of standard stock solution:

The preparation of standard stock solution of carvedilol was done by taking 100mg of the drug in a 100mL volumetric flask. Then, 100ml of methanol was added to obtain a concentration of 1mg/mL. The standard stock solution was again diluted to get a final concentration of 10μg/mL.

 

Preparation of sample solution:

The preparation of sample solution was done by taking twenty tablets of carvedilol which were powdered in a mortar pestle and then, an amount of the powdered sample equivalent to 100mg of drug was taken in a 100 mL volumetric flask and solubilized with 100mL methanol to produce a solution of 1mg/mL. This solution was filtered and diluted to get working concentration in the range of 3-5μg/mL

 

Determination of λmax:

For the determination of λmax, the standard stock solution of carvedilol was diluted with methanol to produce a concentration of 10μg/mL. This solution was scanned in the UV region from 200-400nm to obtain a spectrum. Then, as the λmax was obtained at 242nm, the other wavelengths were taken above and below the λmax fixing 242nm to be the center wavelength. [Figure 2]

 

 

METHOD VALIDATION41:

The validation of the developed method was done according to the ICH Q2 (R1) guidelines which included the validation of parameters like linearity, range, accuracy precision and the limits of quantitation and detection.

 

Linearity:

The first step in the assessment of linearity is to set the concentration range to be studied which is 80% - 120% of the optimum concentration to obtain the absorbance value between 0.2 - 0.9. The assessment of linearity was done by selecting a concentration range of 3-5μg/mL. The absorbance of the aliquots were measured over a range of wavelength surrounding 242nm i.e., 238, 240, 242, 242 and 244nm. The absorbance was recorded as shown in Figure:3 and the calibration plot was obtained.

 

Sensitivity:

Sensitivity of the developed method was obtained by the quantification and detection limits of the method. The limit of detection (LOD) was calculated as:

 

LOD = (3.3 σ) / S …………………………………... (9)

and

The limit of quantitation (LOQ) was calculated as:

 

LOQ = (10 σ) / S ………………………………….. (10)

Where, σ → Standard deviation of the response and S → Slope of the calibration curve.

 

Precision:

Precision of the method was checked by repeating the process i.e., scanning the same concentration and evaluating the deviations in the absorbance. In this method, 4μg/mL concentration is evaluated for precision. The process was repeated three times a day to assess intra-day precision and for three days to assess inter-day precision. The absorbance values at five wavelengths were recorded and the deviation was evaluated. [Figure 4 and 5]

 

Accuracy/Recovery:

There are different approaches to check the accuracy of a method. Standard addition method is one of them. In standard addition method, standard solution is added to the sample solution and scanned. The absorbance value is compared with the standard absorption value to find the recovery or the accuracy of the method. 10mL of solution was pipetted out each from the prepared stock solutions of standard and sample. This was diluted to 100mL to produce a concentration of 100μg/mL. From the diluted standard solution, 0.2mL was taken into 3 different standard flasks of 10mL, then 0.1mL, 0.2mL and 0.3mL of diluted sample solution was added to these flasks and the volume was produced to 10mL with methanol. This resulted in aliquots of three concentrations 3μg/mL, 4μg/mL, and 5μg/mL. Three replicates of each were made and scanned using UV spectrophotometer under 200-400 nm and the % recovery was calculated at all the five wavelengths. [Figure 6]


 


RESULTS AND DISCUSSION:

The absorption maxima (λmax) of carvedilol in a solution of methanol was found to be 242nm. The concentration range selected as per ICH Q2 (R1) guidelines for analysis was 3 – 5μg/mL. The data from the scanned spectrum were presented in calibration graphs which show that they are linear for the selected concentration range. The linear regression equation was found to be: y = 0.1167 x + 0.01944. The correlation coefficient of the linear regression analysis data from the calibration plots showed good linear relationship with R2 = 0.9989. The % RSD for precision of intra-day was found to be 0.71 - 1.09, and inter-day was found to be 1.52 – 1.75. The limit of detection was 0.036μg/mL and the limit of quantitation was 0.109μg/mL, which were within the acceptance limits as specified in ICH Q2 (R1) guidelines.

Linearity:

The absorbance for five concentrations 3, 3.5, 4, 4.5, and 5μg/mL were recorded at 238, 240, 242, 244 and 246 nm [Table 1], and their calibration graphs along with their residual plots are shown in the Figures [Figure 7-16]. The value of SD at all wavelengths is low which indicates that the method has a good linear relationship between the analyte concentration and the absorbance.

 

Table 1: Absorbance at five wavelengths (multivariate)

Concentration (μg/mL)

Absorbance

 

238 nm

240 nm

242 nm

244 nm

246 nm

3

0.368

0.387

0.407

0.381

0.296

3.5

0.427

0.449

0.472

0.443

0.345

4

0.49

0.515

0.54

0.507

0.395

4.5

0.548

0.576

0.605

0.568

0.443

5

0.599

0.629

0.661

0.622

0.485


Table 2: Linearity regression data to demonstrate method validity

Wavelength (nm)

Regression equation

Slope

Intercept

R2

LOD (μg/mL)

LOQ (μg/mL)

238

y = 0.1166x + 0.020

0.1166

0.02

0.9987

0.061981

0.18782

240

y = 0.1222x + 0.0224

0.1222

0.0224

0.9986

0.035917

0.10884

242

y = 0.1282x + 0.0242

0.1282

0.0242

0.9989

0.040156

0.12168

244

y = 0.1214x + 0.0186

0.1214

0.0186

0.9991

0.039415

0.11944

246

y = 0.0952x + 0.012

0.0952

0.012

0.999

0.080074

0.24265

 

Table 3: Intra-day absorbance values and statistics for a single concentration

Concentration (μg/mL)

No. of Repetitions

 

 

Absorbance

 

 

 

 

238 nm

240 nm

242 nm

244 nm

246 nm

 

1

0.493

0.519

0.544

0.514

0.393

 

2

0.489

0.515

0.539

0.511

0.389

4

3

0.496

0.52

0.547

0.516

0.391

 

4

0.497

0.521

0.549

0.518

0.394

 

5

0.499

0.523

0.55

0.519

0.397

 

6

0.502

0.527

0.552

0.521

0.401

Mean

 

0.496

0.5208

0.5468

0.5165

0.3942

S. D.

 

0.004561

0.004021

0.004708

0.003619

0.004309

% R.S.D.

 

0.9195

0.7720

0.8610

0.7008

1.0932

 

Table 4: Inter-day absorbance values and statistics for a single concentration

Concentration (μg/mL)

No. of Repetitions

 

 

Absorbance

 

 

 

 

238 nm

240 nm

242 nm

244 nm

246 nm

 

1

0.496

0.52

0.548

0.516

0.396

 

2

0.491

0.515

0.543

0.512

0.391

4

3

0.505

0.529

0.557

0.525

0.403

 

4

0.485

0.509

0.536

0.505

0.386

 

5

0.505

0.53

0.558

0.527

0.404

 

6

0.493

0.519

0.547

0.515

0.394

Mean

 

0.4958

0.5203

0.5482

0.5167

0.3957

S. D.

 

0.00796

0.008091

0.008377

0.008214

0.006947

% R.S.D.

 

1.6054

1.5550

1.5281

1.5898

1.7559

 

Table 5: Data of Recovery studies

Wavelength (nm)

Concentration levels (%)

Final concentration (μg/mL)

Standard amount present (μg/mL)

Sample amount added (μg/mL)

Absorbance

Amount recovered (μg/mL)

% Recovery

 

80

3

2

1

0.3707

3.02

100.72

238

100

4

2

2

0.4843

3.95

98.84

 

120

5

2

3

0.5933

4.95

99.05

 

80

3

2

1

0.3907

3.02

100.95

240

100

4

2

2

0.5100

3.96

99.03

 

120

5

2

3

0.6247

4.96

99.31

 

80

3

2

1

0.4133

3.04

101.56

242

100

4

2

2

0.5397

3.99

99.94

 

120

5

2

3

0.6607

4.99

99.95

 

80

3

2

1

0.3833

3.01

100.61

244

100

4

2

2

0.5010

3.95

98.82

 

120

5

2

3

0.6143

4.94

98.77

 

80

3

2

1

0.2877

2.92

97.18

246

100

4

2

2

0.3760

3.81

95.19

 

120

5

2

3

0.4620

4.76

95.26

 


Precision:

The calculated value of SD for precision was low which proves the method to be precise and suitable for analysis. % RSD for precision was found to be within the range of 0.71 - 1.75. This is well within acceptance criteria of less than 2% at all the five wavelengths. [Table 3 and 4]

 

Recovery:

The percentage recovery of drug from the mixture of standard and sample was found to be in the range of 95.19–101.56% w/w. This is well within the acceptance limit of 95 - 105% w/w as per ICH Q2 (R1) guidelines. [Table 5]

 

CONCLUSION:

This newly developed method of multivariate calibration technique using UV spectrophotometry was validated by the evaluation of validation parameters as per ICH Q2 (R1) guidelines. The study concludes that this method can be applied to the analysis of carvedilol, both bulk as well as pharmaceutical dosage form in industry as a quality control process or even in the routine analysis because this method is cheap, precise, and non-destructive.

 

List of symbols/abbreviations:

nm = Nanometer

μg/mL = Microgram per millilitre

ICH = International Conference on Harmonization

UV = Ultraviolet

g/mol = Gram per Mole

HPLC = High Performance Liquid Chromatography

HPTLC = High Performance Thin Layer Chromatography

M = Molar

NaOH = Sodium hydroxide

MI = myocardial infraction

IHD = ischemic heart disease

 

CONFLICTS OF INTEREST:

Authors declare NO conflicts of interest

 

ACKNOWLEDGEMENT:

The authors are thankful to the Chancellor, SRM Institute of Science and Technology, and the management of SRM College of Pharmacy, SRM Institute of Science and Technology, Kattankulathur for providing various reprographic sources for carrying out this work.

 

REFERENCES:

1.      Karle CA, Kreye VAW, Thomas D, Rockl K, Kathofer S, Zhang W, Kiehnet J. Antiarrhythmic Drug Carvedilol Inhibits HERG Potassium Channels. Cardiovascular Research. 2001; 49: 361-370.

2.      Chen JJ, Shih NL. Carvedilol: An Anti-Inflammatory and Anti-Atherogenic β-blocker. Acta Cardiologica Sinica. 2003; 19: 93-94.

3.      Franciosa JA, Massie BM, Lukas MA, Nelson JJ, Lottes S, Abraham WT, Fowler M, Gilbert EM, Greenberg B. β-blocker Therapy for Heart Failure Outside the Clinical Trial Setting: Findings of a Community-based Registry. American Heart Journal. 2004; 148: 718-726.

4.      Stoschitzky K, Koshucharova G, Lercher P, Maier R, Sakotnik A, Klein W, Liebmann PM, Lindner W. Stereoselective Effects of (R)- and (S)-carvedilol in Humans. Chirality. 2001; 13: 342-346.

5.      Nath B, Nath LK, Kumar PY. Preparation and In vitro Evaluation of Buccoadhesive Tablets of Carvedilol Using Dried Mucilage Powder of Aegle marmelose. Research Journal of Pharmacy and Technology. 2009; 2(4): 762-767.

6.      Vanderhoff BT, Ruppel HM, Amsterdam PB. Carvedilol: The new role of Beta Blockers in Congestive Heart Failure. American Family Physician. 1998; 58(7): 1627-1639.

7.      Ruffolo RR, Gellai M, Heible JP, Willette RN, Niicholas AJ. The Pharmacology of Carvedilol. European Journal of Clinical Pharmacology. 1990; 38: 82-88.

8.      Abshangen U. A New Molecule with Vasodilating and β-Adrenoceptor Blocking Properties. Journal of Cardiovascular Pharmacology. 1987; 10: 23-32.

9.      Streim K, Sponer G, Muller B, Bartsch W. Pharmacological Profile of Caevedilol: A Compound with β-Blocking and Vasodilating Properties. Journal of Cardiovascular Pharmacology. 1987; 10: 33-41.

10.   Bartsch W, Sponer G, Muller B, Kling L, Bohm E, Martin U, Borbe HO. Pharmacological Characteristics of the Stereoisomers of Carvedilol. European Journal of Clinical Pharmacology. 1990; 38: 104-107.

11.   Sponer G, Strein K, Muller B, Bartsch W. Studies on the Mode of Vasodilating Action of Carvedilol. Journal of Cardiovascular Pharmacology. 1987; 10: 42-48.

12.   Schwarz ER, Kersting PH, Reffelmann T, Meven DA, Al-Dashti R, Skobel EC, Klosterhalfen B, Hanrath P. Cardioprotection by Carvedilol. Journal of Cardiovascular Pharmacology. 2003; 8: 207-215.

13.   Nimbalkar VV, Pansare PM, Nawale DS, Nishane BB. Investigation of Carvedilol as an Antidenaturing and Anticataract Agent. Research Journal of Pharmacy and Technology. 2015; 8(5): 511-516.

14.   Desai DC, Karkhanis VV. Simple spectrophotometric estimation of carvedilol phosphate in bulk and tablet dosage forms. International Research Journal of Pharmacy. 2012; 3(2): 114-116.

15.   Patel SA, Patel NJ. Validated Spectrophotometric Methods for the Determination of Carvedilol in Tablets. International Research Journal of Pharmacy. 2011; 2(7): 171-175.

16.   Theivarasu c, Santanu G, Indumarthi T. UV Spectrophotometric Determination of Carvedilol in Pharmaceutical Formulations. Asian Journal of Pharmaceutical and Clinical Research. 2010; 3(4): 64-68.

17.   Shinkar DM, Dhake AS, Setty CM. Development of UV Spectrophotometric Method for estimation of Carvedilol in Bulk and Pharmaceutical Formulations. Asian Journal of Research Chemistry. 2013; 6(10): 956-959.

18.   Vinod T, Sachin G, Omprakash B. Development, Validation and Stability Study of UV Spectrophotometric Method for Determination of Carvedilol Phosphate in Bulk and Pharmaceutical Dosage Forms. International Journal of ChemTech Research. 2017; 10(9): 252-259.

19.   Divya NS, Narayana B. Simple Methods for the Spectrophotometric Determination of Carvedilol. International Scholarly Research Network. 2012; 373215: 6 pp.

20.   Audumbar M, Vikas K, Rohan D, Sachin G, Ritesh B. Simultaneous UV Spectrophotometric Methods for Estimation of Carvedilol and Hydrochlorothiazide in Bulk and Tablet Dosage Form. Asian Journal of Pharmacy and Technology. 2016; 6(1): 15-20.

21.   Rajan VR, Prathamesh PT. UV Spectrophotometric Estimation of Carvedilol Hydrochloride by Second Order Derivative Methods in Bulk and Pharmaceutical Dosage Form. Research Journal of Pharmacy and Technology. 2014; 7(12): 1459-1462.

22.   Sayed MD, Mahmoud AO, Mohamed A-K A-L, Ahmed IH. Development and Validation of a New Spectrofluorometric Method for the Determination of Some Beta-blockers through Fluorescence Quenching of Eosin Y. Open Chemistry Journal. 2016; 14: 258-266.

23.   Sovan LP, Utpal JG, Mohanta P, Manna PK. Method Development and Validation of Reverse Phase High Performance Liquid Chromatography Method to Determine Carvedilol in Pharmaceutical Formulations. Der Pharmacia Sinica. 2013; 4(6): 22-27.

24.   Sivakamasundari G, Kannappan N. Development and Validation of RP-UPLC Method for the Simultaneous estimation of Carvedilol and Hydrochlorothiazide in Pharmaceutical Dosage Form. International Research Journal of Pharmaceuticals. 2018; 9(2): 23-26.

25.   Rao UVM, Haritha G, Krishnaiah T, Saikiran TG, Reddy TA, Nikhil UVS, Veerabomma HS, Veeram SDR. Method Development and Forced Degradation Studies of Carvedilol by RP-HPLC. International Journal of Pharmacy and Analytical Research. 2017; 6(3): 457-468.

26.   Patel H, Jivani N. Development of Validated RP-HPLC Method for Simultaneous Estimation of Carvedilol and Ivabradine. World Journal of Pharmacy and Pharmaceutical Sciences. 2015; 4(5): 630-639.

27.   El-Adl SM, Abdel-Aziz LM, Mohamed MAM. HPLC Determination of Carvedilol and Atorvastatin Calcium in Their Bulk and Dosage Form. Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry. 2016; 4(2): 112-123

28.   Suganthi A, Ravi TK. Development of RP-HPLC Method for the Estimation of Nebivolol and Carvedilol with selected NSAIDS and its Application to Drug Displacement Interaction Studies. Malaysian Journal of Pharmaceutical Sciences. 2014; 12(1): 47-65.

29.   Mahaveer S, Kashkhedikar SG, Soni L, Abhinav G, Tripti G, Amrish P. Development of RP-HPLC Method for Estimation of Carvedilol in Tablet Formulations. Research Journal of Pharmacy and Technology. 2008; 1(1): 18-21.

30.   Clohs L, McErlane KM. Development of a Capillary Electrophoresis Assay for the Determination of Carvedilol Enantiomers in Serum using Cyclodextrins. Journal of Pharmaceutical and Biomedical Analysis. 2001; 24(4): 545-554.

31.   Samin H, Somaieh S, Abolghasem J. A dispersive liquid – liquid microextraction and chiral separation of Carvedilol in Human plasma using Capillary Electrophoresis. Bioanalysis. 2015; 7(9): 1107-1117.

32.   Patel LJ, Suhagia BN, Shah PB. RP-HPLC and HPTLC Methods for the Estimation of Carvedilol in Bulk Drug and Pharmaceutical Formulations. Indian Journal of Pharmaceutical Sciences. 2006; 68(6): 790-793.

33.   Danzer K, Currie LA. Guidelines for Calibration in Analytical Chemistry. Pure and Applied Chemistry. 1998; 70(4): 993-1014.

34.   Arayne MS, Najma S, Zuberi MH, Siddiqui FA. Spectrophotometric Quantification of Metformin in Bulk Drug and Pharmaceutical Formulations using Multivariate Technique. Indian Journal of Pharmaceutical Science. 2009; 71(3): 331-335.

35.   Susmitha S, Kokilambigai KS, Lakshmi KS. Spectrophotometric Quantification of Telmisartan Employing Multivariate Calibration Technique in Bulk and Pharmaceutical Formulations. Research Journal of Pharmacy and Technology. 2019; 12(4): 1799-1805

36.   Kokilambigai KS, Seetharaman R, Kavitha J, Sowndaravel P, Lakshmi KS. Multivariate Calibration Technique for the Spectrophotometric Quantification of Zaleplon in Bulk Drug and Pharmaceutical Formulations. Journal of Pharmaceutical Sciences and Research. 2017; 9(6): 824-829.

37.   Madhan S, Kavitha J, Lakshmi KS. Multivariate Calibration Technique for the Spectrophotometric Quantification of Ivermectin in Pharmaceutical Formulation. Asian Journal of Pharmaceutical and Clinical Research. 2019; 12(2): 444-451.

38.   Santosh VG, Parag LK, Atul PC. Application of Multivariate Calibration Methods for Simultaneous Determination of Drugs in Fixed Dose Combination. Asian Journal of Pharmaceutical Analysis. 2018; 8(1): 1-6.

39.   Kathirvel S, Sahiti V, Suneetha A. Spectrophotometric Estimation of Risedronate Sodium in Bulk and Pharmaceutical Formulations using Multivariate Technique. Research Journal of Pharmacy and Technology. 2010; 3 (4): 1209-1212.

40.   Abdul SM, Swetha D, Nikhat F, Humera B, Syeda SS, Mohammad AS, Nuha R. An Overview of Validation and Basic Concepts of Process Validation: Quality Assurance View Point. Asian Journal of Pharmacy and Technology. 2016; 6 (3): 169-176.

41.   Q2R1 ICH guidelines for analytical method development. Available at:http://www.ich.org/fileadmin/Public_Web_Site/ICH_Products/Guidelines/Quality/Q2_R1/Step4/Q2 _R1__Guideline.pdf.

 

 

 

Received on 16.07.2019            Modified on 21.08.2019

Accepted on 30.09.2019           © RJPT All right reserved

Research J. Pharm. and Tech 2020; 13(2):.915-922.

DOI: 10.5958/0974-360X.2020.00173.0